Classification of sphygmomanometers/blood pressure measuring devices A table of blood pressure measurement devices as recommended by the dabl Educational Trust Advisory Board Table of devices currently available on the market Table of devices that have been discontinued by their manufacturers
Table of currently available blood pressure monitors Devices used in a clinical setting Table of devices used for self/home blood pressure measurement Table of devices used for Ambulatory Blood Pressure Measurement

Sphygmomanometers for Ambulatory Blood Pressure Measurement

The following table is a list of all ABPM devices currently available. Discontinued devices are shown on a separate table. A complete list of all devices is available on our Device Index.

Devices for ABPM

New to the marketNew to the website  
Device Mode AAMI BHS ESH Circumstance Recommendation Ref  
A&D TM-2430 Osc Pass A/A   At rest Recommended 6  
BpLab Osc     Pass At rest Questionable 19  
Cardiette BP One Osc   A/A   At rest Recommended 25
HealthStats BPro Wrist Pass   Pass Study details omitted Not Recommended 26
IEM
Mobil O Graph (version 12)
Osc Pass B/A   At rest Recommended 14  
Jotatec P24B Osc     Pass Abstract Only Questionable 27
Meditech ABPM-04 Osc Pass B/B   At rest Recommended 1  
Meditech ABPM-05 Osc Pass B/B   ABPM-04 Equivalence Recommended E14
Nissei DS-250 Osc     Pass Abstract; 'Passes' incorrectly reported Questionable 17  
Aus     Pass
Save 33, Model 2 Osc Pass B/B   At rest Recommended 23
Seinex SE-25M Osc Pass A/B   Analysis violations (AAMI & BHS); BHS 'pass' incorrectly reported Questionable 16  
Spacelabs 90207 Osc Pass B/B   At rest Recommended 2  
Pass B/B   In pregnancy Recommended 3
Pass A/B   Elderly standing & sitting SBP < 161 mmHg Recommended 4
Pass A/C   In pregnancy Not Recommended 7
Pass B/C   In pregnancy Not Recommended 8
Fail D/D   In pre-eclampsia Not Recommended 9
Pass C/C   In pre-eclampsia Not Recommended 10
Pass/ Fail C/D   In children Not Recommended 11
Pass A/D   Elderly supine over all pressures Not Recommended 12
Pass C/B   During haemodialysis Not Recommended 13
Spacelabs 90217 Osc Pass A/A   At rest Recommended 5  
Suntech Medical AGILIS Osc     Pass At rest Recommended 21  
Suntech Medical OSCAR 2 Osc     Pass At rest; Excellent Paper Recommended 18  
  A/A   Excellent study and paper Recommended 24  
Tensioday Osc Pass A/A   At rest; grades of pressure; varying positions Recommended 15  
Tonoport V Osc     Fail At rest Not Recommended 20  
    Pass At rest Questionable  22  

References for Validation of Devices for ABPM

1 Barna I, Keszei A, Dunai A. Evaluation of Meditech ABPM-04 ambulatory blood pressure measuring device according to the British Hypertension Society protocol. Blood Press Monit. 1998;3:363-368
2 O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the Spacelabs 90207 determined by to the British Hypertension Society Protocol. J Hypertens 1991;9(suppl 5):S25-S31
3 Shennan AH, Kissane J, de Swiet M. Validation of the Spacelabs 90207 ambulatory blood pressure monitoring in pregnancy. Br J Obstet & Gynae 1993;100:904-8
4 Iqbal P, Fotherby MD, Potter JF. Validation of the Spacelabs 90207 automatic non-invasive blood pressure monitor in elderly subjects. Blood Press Monit 1996;1:367-73
5 Baumgart P, Kamp J. Accuracy of the Spacelabs Medical 90217 ambulatory blood pressure monitor. Blood Press Monit 1998;3:303-307
6 Palatini P, Frigo G, Bertolo O, Roman E, Da Corta R, Winnicki M. Validation of the A&D TM-2430 device for ambulatory blood pressure monitoring and evaluation of performance according to subjects’ characteristics. Blood Press Monit 1998;3:255-260
7 O'Brien E, Mee F, Atkins N, Halligan A, O'Malley K. Accuracy of the Spacelabs 90207 ambulatory blood pressure measuring system in normotensive pregnant women determined by the British Hypertension Society Protocol. J Hypertens 1993;11(suppl 5):S282-S283
8 Franx A, van der Post JAM, van Montfrans GA, Bruinse HW. Comparison of an auscultatory versus an oscillometric ambulatory blood pressure monitor in normotensive, hypertensive and preeclamptic pregnancy. Hypertension in pregnancy 1997;16:187-202
9 Natarajan P, Shennan AH, Penny J, Halligan AW, de Swiet M, Anthony J. Comparison of auscultatory and oscillometric automated blood pressure monitors in the setting of preeclampsia. Am J Obstet Gynecol 1999;181:1203-1210
10 Shennan A, Haligan A, Gupta M, Taylor D, de Swiet M. Oscillometric blood pressure measurements in severe preeclampsia: validation of the Spacelabs 90207. Br J obstet & Gynaecol 1996;103:171-3
11 Belsha CW, Wells TG, Rice HB, Neaville WA, Berry PL. Accuracy of the Spacelabs 90207 ambulatory blood pressure monitor in children and adolescents. Blood Press Monit 1996;1:127-133
12 Iqbal P, Fotherby MD, Potter JF. Validation of the Spacelabs 90207 automatic non-invasive blood pressure monitor in elderly subjects. Blood Press Monit 1996;1:367-73
13 Peixoto AJ, Gray TA, Crowley ST. Validation of the Spacelabs 90207 ambulatory blood pressure device for hemodialysis patients. Blood Press Monit 1999;4:217-221
14 Jones CR, Taylor K, Chowienczyk P, Poston L, Shennan A. A validation of the Mobil O Graph (version 12) ambulatory blood pressure monitor. Blood Press Monit 2000;5:233-238
15 Nemeth Z, Moczar K, Deak G. Evaluation of the Tensioday ambulatory blood pressure monitor according to the the protocols of the British Hypertension society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 2002;7:191-197
16 Tannahill G, Gilmore C, Silke B. Validation of the upper arm algorithm for blood pressure determination incorporated in the Seinex digital devices using the AAMI and British Hypertension Society protocols Northern Ireland Medicine Today 2002;33:26-27
17 Altunkan Ş, Iliman N, Altunkan E. Validation of the Nissei DS-250 ambulatory blood pressure monitoring device according to the International Protocol J Hypertens 2003;[dabl® Educational Trust regrets that full details are missing]
18 Jones, SC, Bilous M, Winship S, Finn P, Goodwin J. Validation of the OSCAR 2 oscillometric 24-hour ambulatory blood pressure monitor according to the International Protocol for the validation of blood pressure measuring devices. Blood Press Monit 2004;9:219-223
19 Rogoza AN. Accuracy of blood pressure estimation by automated monitors Functional diagnostics 2003;1:56-64. (In Russian)
20 O'Brien E, Atkins N, Murphy A, Lyons S. Validation of the Tonoport V ambulatory blood pressure monitor according to the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults Blood Press Monit 2003;8:255-260
21 Mallion JJM, Pierre H, Neuder Y, Ormezzano O, Baguet JP Validation of the AGILIS ambulatory blood pressure monitor according to the European society of Hypertension International Protocol for Validation of Blood Pressure Measuring devices in Adults Blood Press Monit 2005;10:97-101
22 Haensel A, Utech K, Langewitz W Validation of the TONOPORT V blood pressure measuring monitor in adults Journal of Human Hypertension 2005; 19 :745-749
23 De Gaudemaris R, White A, Pascal I, Siche JP, Baguet JP, Mallion JM. Évaluation du moniteur de pression artérielle en ambulatoire Save 33 modèle2. Arch Mal Coeur 1999;92:1133-1137
24 Goodwin J, Bilous M, Winship S, Finn P, Jones SC. Validation of the Oscar 2 oscillometric 24-h ambulatory blood pressure monitor according to the British Hypertension Society protocol Blood Pressure Monitoring 2007;12:113-117
25 Friz HP, Sega R, Facchetti R, Primit L, Beltrame L, Bombelli M. Accuracy evaluation of the 'Cardiette BP one' ambulatory blood pressure monitor Blood Pressure Monitoring 2008;13:107-110
26 Nair D, Tan SY, Gan HW, Lim SF, Tan J, Zhu M, Gao H, Chua NH, Peh WL, Mak KH. The use of ambulatory tonometric radial arterial wave capture to measure ambulatory blood pressure: the validation of a novel wrist-bound device in adults Journal of Human Hypertension 2008;22:220-222 (Research Letter)
27 Octavio JA, Leal N, Contreras J, Octavio B, Cressa L, Valecillos I, Castellanos D, Fabiano D. Accuracy of Jotatec®-P24B device for teletransmitted ambulatory blood pressure monitoring Journal of Hypertension 2008;26 (suppl. 1):S493 (Abstract PS34/MON/60)

© Copyright dabl®Educational Trust Ltd 2008. All rights reserved. Please read Disclaimer before using this website.