| Device |
Mode |
AAMI |
BHS |
ESH |
Circumstance |
Recommendation |
| A&D TM-2564G |
Osc |
Pass |
A/A |
|
Serious protocol violations |
Questionable |
BpTRU BPM-100 BPM-300 |
Osc |
Pass |
A/A |
|
At rest. Pressure ranges |
Recommended |
| Pass |
|
|
Children |
Questionable |
| Pass |
A/A |
|
Mixed adult/paediatric population Protocol Violations Should have used 1990 instead of 1993 BHS criteria but still A/A |
Questionable |
BpTRU (unspecified) |
|
|
|
|
Pass Ad Hoc Protocol; Abstract |
Questionable |
| Colin 8800C |
Osc |
Pass |
|
|
AAMI only |
Questionable |
| Colin Pressmate BP10 |
Osc |
|
|
Pass |
Abstract only |
Questionable |
| CVProfilor DO-2020 |
Dynamic linear deflation, Osc & Acoustic |
|
|
|
New methodology, Passed Ad-hoc protocol |
Questionable |
| Datascope Accutorr Plus |
Osc |
Pass |
A/A |
|
At rest |
Recommended |
| Pass |
|
|
Trimline cuff replacement, at rest, protocol variation |
Questionable |
| |
|
Pass |
Children. Adaptation of protocol along BHS guidelines. |
Recommended |
DiaTecne PulsePen |
Tonometry |
|
|
|
Abstract; Pass ad hoc protocol; Developer assessment |
Questionable |
| Dinamap 1846 XT |
Osc |
Pass |
C/C |
|
At rest; varying pressures, Recruitment violation, procedure violation |
Questionable |
| Dinamap ProCare |
Osc |
|
|
Pass |
At rest |
Recommended |
| Osc |
Pass |
A/A |
|
At rest |
Recommended |
| Dinamap ProCare |
Osc |
|
|
Pass |
At rest |
Recommended |
G-Care Electronics SP-800 |
Osc |
|
|
Pass |
Abstract Only |
Questionable |
| HoMedics BPA-300 |
Aus & Osc |
Pass |
|
|
Sample size well over 85. Low mean BP |
Questionable |
| Lohmeier B-606 |
Osc |
|
|
Pass |
Abstract |
Questionable |
| Microlife WatchBP O3 3MZ0 |
Osc |
|
|
Pass |
At rest |
Recommended |
| Microlife WatchBP Office |
Osc |
|
|
Pass |
At rest |
Recommended |
Omron 705IT (See also SBPM) |
Osc |
|
|
Pass |
|
Recommended |
| Pass |
A/A |
|
Minor recruitment modification |
Recommended |
| |
|
Pass |
Children and adolescents. Reasonable adaptation of protocol. |
Recommended |
| Omron 725 CIC |
Osc |
|
|
|
Ad Hoc Protocol. Pointless study |
None |
| Omron 907 |
Osc |
|
|
Pass |
Data analysed according to IP proposal only |
Questionable |
| Pass |
|
|
At rest; AAMI only, Recruitment violation |
Questionable |
| Osc & Aus |
|
|
Pass |
Elderly Population. Mode not stated. |
Recommended |
| Omron M4-I |
Osc |
|
|
Pass |
705IT Equivalence |
Recommended |
| Primo RACD |
Osc |
(Pass) |
|
|
RAL catheter in ICU. Limited sample. Device "shows promise" (Authors). |
Questionable |
| Spengler Tensionic Mod EPS 112 |
Osc |
Pass |
B/A |
|
At rest; Abstract |
Questionable |
| Tensoval duo control |
Aus & Osc |
Pass |
A/A |
Pass |
Auscultatory validation only |
Recommended |
| Terumo ES-H55 |
Osc Double-cuff |
Pass |
|
|
Manufacturer Assessment |
Questionable |
| Welch Allyn Spot Vital Signs |
Osc |
Pass? |
|
|
Only Method 1 completed. Ages 5-77 |
None |
| Welch Allyn 'Vital Signs' |
Osc |
Pass |
A/A |
|
At rest. Pressure ranges |
Recommended |
| Fail |
C/D |
|
At rest |
Not Recommended |
| |
|
Fail |
Children. Adaptation of protocol along BHS guidelines. |
Not Recommended |